Diaphragmatic excursion and extubation success

Diaphragmatic excursion and extubation success

Knowing when to extubate an ELBW is never an easy task. Much has been written about extubation checklists including such measures as mean airway pressure minimums and oxygen thresholds as well as trials of pressure support at low rates. The fact remains that no matter how hard we try there are those that fail even when all conditions seem to be met for success. The main culprit has been thought to be weakening of the diaphragm as the infant stays on the ventilator for longer periods of time. Specifically, myofibrillar contractile dysfunction
and myofilament protein loss are what is occurring leading to a weakened diaphragm which may be incapable of supporting the infant when extubated even to CPAP. More recently in Neonatology the use of point of care ultrasound (POCUS) has gained in popularity and specifically use of lung ultrasound has helped to better classify various disease conditions not only in determining which disease is active but also following its course. Using POCUS to measure thickness and excursion of the diaphragm has been employed in the adult world so using it in neonates to determine extubation readiness seems like a logical next step.

An Observational Cohort Study

Bahgat E et al published Sonographic evaluation of diaphragmatic thickness and excursion as a predictor for successful extubation in mechanically ventilated preterm infants in the European Journal of Pediatrics. This small study sought to look at preterm infants born under 32 weeks and assessed a number of measurements of their diaphragm bilaterally including thickness of both during the respiratory cycle and the excursion (measured as most caudad and cephalad position during respiration). All patients underwent a similar process prior to extubation using PSV with a support of +4 over peep with measurements taken 1 hour prior to planned extubation. All infants met unit criteria for a trial of extubation based on blood gases, FiO2 and MAP being less than 8 cm H2O. All infants received a PSV trial for 2 hours before being extubated to CPAP +5. The sonographic assessment technique is laid out in the paper and the study end point was no reintubation in the 72 hours after extubation. The decision to reintubate was standardized as follows: more than six episodes of apnea requiring stimulation within 6 h, or more than one significant episode of apnea requiring bag and mask ventilation, respiratory acidosis (PaCO2 > 65 mmHg and pH < 7.25) or FiO2 > 60% to maintain saturation in the target range (90–95%).

Differences between the groups at baseline included a longer median day of extubation by 3 days, total duration of mechanical ventilation, higher mean airway pressure and FiO2 all in in the failure group.

Results of the study find a key difference in measurements

Looking at table 2 below the main finding of the study was that the biggest difference between those infants who succeeded and those that failed was the excursion of the diaphragm rather than the thickness. The greater the excursion the better the chance at successful extubation. In experienced hands the measurement does not take that long to do either.

As the authors point out in the paper:

“A right hemidiaphragmatic excursion of 2.75 mm was associated with 94% sensitivity and 89% specificity in predicting successful extubation. A left hemidiaphragmatic excursion of 2.45mmwas associated with 94% sensitivity and 89% specificity in predicting successful extubation”

Is this the holy grail?

There is no question that this technique adds another piece to the puzzle in helping us determine when it is safe to extubate. If I can pick one fault with the study it is the use of a pressure of +5 to support the extubated infants. If you look at the mean level of MAP the infants were on prior to extubation in the two groups it was 6.3 in the successful group and 6.6 in those who failed. By choosing to extubate the group that was already on a mean of about 24% to an even lower pressure level I can’t help but wonder what the results would have looked like if extubation occurred at a non-invasive level above that when they were intubated. Our unit would typically choose a level of +7 to extubate such infants to and avoid pulmonary volume loss so what would the results show if higher pressures were used (someone feel free to take this on).

One thing though that is borne out of all this however is that if diaphragmatic weakening happens in the neonate with prolonged ventilation as well it would be supported by the long length of ventilation in the failure group that also has less diaphragmatic thickness and excursion. What this study in my mind really says is that extubation should occur as early as possible. Every time you hear someone say “why don’t we wait one more day” you can now imagine that diaphragm getting just a little weaker.

As I said on a “tweet” recently “No one should brag about having a 100% extubation success rate”. If that is your number you are waiting too long to extubate. Based on the information here it should be a reminder that the plan for extubation needs to start as soon as the tube is inserted in the first place.

Point of Care Ultrasound (POCUS)

Welcome to the POCUS section!

We hope to provide education through links to publications and videos demonstrating the benefits of adopting POCUS! Less ionizing radiation and enhanced diagnostic accuracy are just two of the benefits of using such techniques. Videos demonstrating and discussing this technique can be found on the Point of Care Neonatal Ultrasound Playlist on my Youtube channel
Background

Use of point of care ultrasound has expanded over the last decades particularly in intensive care to the point that it is now readily available for use by the clinical care practitioners in this setting (1). Today clinicians are using ultrasound at the bedside to assist in the evaluation of physiological abnormalities in a number of body systems. Ultrasound has been used to image body organs for over 50 years (2). It is currently the most widely used imaging modality in medicine. Advantages are that ultrasound is portable, free of radiation risk and relatively inexpensive compared to other diagnostic modalities like magnetic resonance and computed tomography (3). The main limitation when considering this for use by NICU clinicians is that it requires advanced training. In addition, when compared to traditional x-ray, ultrasound has limited penetration to air and bones and therefore structures deeper to them cannot be well assessed (4).
Indications:
There are three general indications for ultrasound in the neonatal setting:
1) Anatomic assessment of static organs such as the brain, lungs, liver, kidney and spleen to evaluate for anomalies, hemorrhage, space occupying lesions and abnormal fluid collections.
2) Dynamic assessment of moving organs, such as the heart, lungs, intestine, and the vascular system to evaluate blood flow and physiologic processes.
3) Locating vessels for cannulation and determining the position of catheter tips.

Anatomic ultrasound assessment of static organs should be provided by a trained radiologist. Dynamic ultrasound assessment can be performed by a trained neonatal clinician who understands the clinical details of his or her patients and is familiar with the underlying pathophysiologic mechanisms (5). Table 1 shows different applications performed by a trained clinician.
Point of care lung ultrasound:

In the last 10 years, research studies have shown that lung ultrasound (LUS) is an accurate, non-invasive method for predicting ventilatory failure and offers advantages over traditional chest radiography (6). LUS can accurately and reliably diagnose transient tachypnea of the newborn (TTN) and has a great value in differentiating TTN from respiratory distress syndrome (RDS) (7). Additionally, many of the other common pulmonary and pleural diseases in neonates display specific findings on LUS which can be useful in the differential diagnosis (8).
We developed a screening model of bedside lung ultrasound assessment for infants requiring respiratory support5. Like any other diagnostic technique it should be only used in integration with the clinical assessment and interpreted according to the clinical presentation of the individual patient while considering particular limitations of this modality.
Point of care intestinal ultrasound:

Necrotising enterocolitis is a serious disorder in infants and commonly associated with complications like short bowel syndrome and total parenteral nutrition related issues. The reported mortality is up to 40%, so early diagnosis and management are essential (9). The radiographic diagnosis by XR after clinical suspicion is still the standard in most centers. The main issue with radiograph is being limited to 3 main findings, pneumatosis intestinalis (PI), portal vein gases (PVG), and perforation, and radiograph diagnosis of PI and PVG is sometimes a challenge with low sensitivity and wide range of inter-observer variability. There has been increasing evidence that with real-time ultrasound, PI and PVG can be better detected than with x-ray (10). Ultrasound is able to assess the bowel wall directly and detect bowel wall thickening or thinning, reduced peristalsis or abnormal bowel wall perfusion by color Doppler. Peritoneal fluid, both intraluminal and extra luminal is also visible (11,12). This can be performed in any suspected case with compromised intestinal performance like intestinal obstruction or ischemia and not only in cases with suspected NEC.

Table 1: different applications performed by either professional sonographer (radiologist or cardiologist) or a trained clinician

Cranial ultrasound
emergency assessment of suspected hemorrhage
Doppler assessment of cerebral arteries in cases hemodynamic instability e.g. PDA
Intestinal ultrasound Urgent evaluation of suspected necrotizing entercolitis, intestinal ischemia
Lung ultrasound New emerging modality for assessment of common neonatal lung diseases, e.g. RDS, TTN, meconium, pneumothorax, pleural effusions.
Focused heart ultrasound Assessment of specific neonatal hemodynamics issues
Vascular assessment Blood flow by Doppler for assessment of resistance or shunting of blood through arteriovenous malformation or PDA
Interventional POCUS Central line placement, lumber puncture, bladder tapping for urine sample. Peritoneal and pericardial tap of significant effusions

References:
1. Evans N, Gournay V, Cabanas F, et al. Point-of-care ultrasound in the neonatal intensive care unit: international perspectives. Semin Fetal Neonatal Med. 2011;16(1):61-68. doi:10.1016/j.siny.2010.06.005.
2. Pereda M a., Chavez M a., Hooper-Miele CC, et al. Lung ultrasound for the diagnosis of Pneumonia in Children: A Meta-analysis. Pediatrics. 2015;135(4):714-722. doi:10.1542/peds.2014-2833.
3. Escourrou G, De Luca D. Lung ultrasound decreased radiation exposure in preterm infants in a neonatal intensive care unit. Acta Paediatr. 2016:n/a-n/a. doi:10.1111/apa.13369.
4. Volpicelli G, Elbarbary M, Blaivas M, et al. International evidence-based recommendations for point-of-care lung ultrasound. Intensive Care Med. 2012;38(4):577-591. doi:10.1007/s00134-012-2513-4.
5. Elsayed Y, Abdelmawla M, Narvey M. A model of integrated lung and focused heart ultrasound as a new screening examination in infants at risk of respiratory or hemodynamic compromise. 2017;6(1):1-14. doi:10.7363/060131.
6. Xirouchaki N, Magkanas E, Vaporidi K, et al. Lung ultrasound in critically ill patients: comparison with bedside chest radiography. Intensive Care Med. 2011;37(9):1488-1493. doi:10.1007/s00134-011-2317-y.
7. Liu J, Cao H-Y, Wang X-L, Xiao L-J. The significance and the necessity of routinely performing lung ultrasound in the neonatal intensive care units. J Matern Neonatal Med. 2016;7058(March):1-6. doi:10.3109/14767058.2016.1152577.
8. Copetti R, Cattarossi L. Lung Ultrasound in Newborns, Infants, and Children. 2011:241-245. doi:10.1007/978-3-642-21247-5.
9. Dilli D, Suna Oğuz S, Erol R, Ozkan-Ulu H, Dumanlı H, Dilmen U. Does abdominal sonography provide additional information over abdominal plain radiography for diagnosis of necrotizing enterocolitis in neonates? Pediatr Surg Int. 2011;27(3):321-327. doi:10.1007/s00383-010-2737-8.
10. Bohnhorst B. Usefulness of abdominal ultrasound in diagnosing necrotising enterocolitis. Arch Dis Child Fetal Neonatal Ed. 2013;98:F445-50. doi:10.1136/archdischild-2012-302848.
11. Gale HI, Gee MS, Westra SJ, Nimkin K. Abdominal ultrasonography of the pediatric gastrointestinal tract. World J Radiol. 2016;8(7):656. doi:10.4329/wjr.v8.i7.656.
12. Kim H-Y, Kim I-O, Kim WS, Kang GH. Bowel sonography in sepsis with pathological correlation: an experimental study. Pediatr Radiol. 2011;41(2):237-243. doi:10.1007/s00247-010-1806-4.

Point of Care Ultrasound: Changing Practice For The Better in NICU

Point of Care Ultrasound: Changing Practice For The Better in NICU

It has been some time since I wrote on the topic of point of care ultrasound (POC). The first post spoke to the benefits of reducing radiation exposure in the NICU but was truly theoretical and also was really at the start of our experience in the evolving area.  Here we are a year later and much has transpired.

We purchased an ultrasound for the NICU in one of our level III units and now have two more on the way; one for our other level III and one for our level II unit.  The thrust of these acquisitions have been to reduce radiation exposure for one but also to shorten the time to diagnosis for a number of conditions.  No matter how efficient x-ray technologists are, from the time a requisition is placed  to the arrival of the tech, placement of the baby and then processing of the film, it is much longer than using a POC at the bedside.  Having said that though is it accurate?  There are many examples to choose from but when thinking about times when one would like an answer quickly I can’t think of anything much better than a pneumothorax.

Chest X-ray vs POC for Diagnosis of Pneumothorax

The diagnosis of a pneumothorax is easily diagnosed by ultrasound when there is an absence of lung sliding as seen in this video.  In the majority of cases employing POC we are looking at ultrasound artifacts.  In the case of pleural sliding which is best described as ants marching, it’s absence indicates the presence of a pneumothorax.  The “lung point” sign as shown in this video marks the transition from pleural sliding to none and in a mode called “M” appears as a bar code when the pneumothorax is present.

Using such signs Raimondi F et al as part of the LUCI (Lung Ultrasound in the Crashing Infant) group compared traditional x-ray diagnosis as the gold standard to POC for diagnosis of pneumothorax.  This study is important as it demonstrated two very important things in the 42 infants who were enrolled in the study.  The first was the accuracy of POC.  In this study each patient had both an ultrasound and an x-ray and the results compared to determine how accurate the POC was.  Additionally in cases where there was no time for an x-ray to confirm the clinical suspicion the accuracy of the study was determined based on the finding of air with decompression along with abrupt clinical improvement.  In case people are wondering infants as small as 24 weeks were included in the study with an average weight of 1531 +/-832 g for included infants.

The accuracy was stunning with a sensitivity and specificity of 100% each.  Comparing this with clinical evaluation (transillumination, assessment of breath sounds) was far less accurate with a sensitivity of 84% (65-96) and specificity 56% (30-80).

Adding to the accuracy of the test is the efficiency of the procedure. “After clinical decompensation, lung ultrasound scans were completed in a mean time of 5.3 +/- 5.6 minutes vs a mean time of 19 +/- 11.7 minutes required for a chest radiograph (P < .001).” In short, it is very accurate and can be done quickly.  In an emergency, can you think of a better test?

If efficiency weren’t enough what about the reduction in radiation exposure?

This was the focus of a recent paper by Escourrou G & Deluca D entitled Lung ultrasound decreased radiation exposure in preterm infants in a neonatal intensive care unit. The authors in this study chose to examine retrospecitively the period from 2012 – 2014 as in 2013 they rolled out a program of teaching POC ultrasound to clinicians.  The purpose of this paper was to see if practitioners educated in interpretation of ultrasound would actually change their practice and use less ionizing radiation.

Their main findings are indicated in the table

Test 2012 2014 p
Min 1 x-ray during admission 81% 70% <0.001
Total x-rays 1976 1476
Mean x-rays per patient 4.9+/-1.5 2.6+/-1.0 <0.001
Mean radiation dose (microGy) 183+/-78 68+/-30 <0.001

As they predicted use of ionizing radiation dropped dramatically.  I should also mention that they tracked outcomes such as IVH, mortality and BPD to name a few and found no change over time.  In conclusion the use of ultrasound did not affect major outcomes but did spare each neonate ionizing radiation.

Now before anyone hits the panic button I still think the amounts of radiation here are safe for the most part.  In Canada the maximum allowed dose for the public per year is 1 mSv which is the equivalent of 1000 microGy.  This was obtained from the Nuclear Safety agency in Canada in case you are interested in finding out more about radiation safety limits.

Back in 2012 at least in this study, 2 standard deviations from the mean would have put the level received at a little over a third of what the annual limit is but it is the outliers we need to think of.  What about kids getting near daily x-rays while on high frequency ventilation or for monitoring pleural fluid collections?  There certainly are many who could receive much higher dosages and it is for those kids that I believe this technology is so imperative to embrace.

It will take time to adopt and much patience.  With any new roll out there is a learning curve.  Yes there will be learners who will need to handle patients and yes there will be studies done at times to obtain the skills necessary to perform studies in an efficient and correct manner but I assure you it will be worth it.  If we have a way of obtaining faster and accurate diagnoses and avoiding ionizing radiation don’t we owe it to our patients and families to obtain such skill?  I look forward to achieving a centre of excellence utilizing such strategies and much like this last study it will be interesting to look back in a year an see how things have changed.

 

 

Point of Care Ultrasound in the NICU

Point of Care Ultrasound in the NICU

This post is meant to supplement an earlier post on the same topic.

Since introducing POC U/S in our unit there has been great enthusiasm and we will begin shortly introducing our nursing group to it’s use in order to enhance usage.  Now that we have had some exposure we took the time to capture some our thoughts on this technology in this accompanying video graciously supported by the Children’s Hospital Foundation in Winnipeg.  To watch this wonderful video by Dr. Ganesh Srinivasan a Neonatologist and technology aficionado in our institution click on the link below.

Reducing Radiation Exposure in Neonates:  Replacing Radiographs With Bedside Ultrasound.

Reducing Radiation Exposure in Neonates: Replacing Radiographs With Bedside Ultrasound.

Zonare

To quote Malcolm Gladwell, I think I witnessed a “Tipping Point” yesterday.   The standard of practice for determining the presence or absence of pneumonia and pneumothorax has always been a chest radiograph. Determination of fluid collections by ultrasonography in the chest or abdomen, the domain of the Pediatric Radiologist.

Yesterday however, I was introduced to the use of bedside Point of Care Ultrasound (POC U/S) for the diagnosis of all of these aforementioned conditions. I spent the morning at a course on POC U/S for the Neonatologist and with that became comfortable much sooner than I expected with making such diagnoses based on some very straightforward criteria. Now before you think I am getting ahead of myself, I do not believe for a second that I am now competent to use U/S to replace radiographs as the “gold standard” in my practice. It has led me to consider though how one might go about reaching this level of confidence but I will share more on that later.

An example of how one can use ultrasound to exclude a pneumothorax is shown here https://www.youtube.com/watch?v=ws7DI4ZGQCo. The presence of bilateral “lung sliding” of the parietal and visceral pleura accompanied by “comet trails” excludes a pneumothorax.   The absence of these findings is suggestive of a pneumothorax although other pathological findings can be present necessitating a chest x-ray. Think of how many times a patient develops a sudden increase in FiO2 and you do a chest x-ray to exclude a pneumothorax before increasing PEEP. This could change with acquisition of such technology.

The first question to really ask before taking the leap to replace or augment the classic chest radiograph is to look at the literature. With respect to the Neonate, the body of literature is not large but there are a couple of recent papers that are worth mentioning.   The first is by Pereda MA et al. Lung Ultrasound for the Diagnosis of Pneumonia in Children: A Meta-analysis. (http://1.usa.gov/1CM24Sv) The review examined 9 studies of which 2 were in the neonatal population. The findings were intriguing. The sensitivity was 96% and specificity 93% when compared to the “Gold Standard” of either CXR or CXR with clinical diagnosis. To put this into context, there were 765 children who would have had thousands of x-rays during their hospital stay. This technology could reduce the number to ionizing radiation exposure by an extraordinary amount given the accuracy of the test compared to standard x-rays. The benefit of such reduction can be immediately appreciated when one considers that children are at an increased risk of malignancy compared to adults from ionizing radiation.  Free supporting publication here.

The second paper to highlight is by Lovrenski J. Lung Ultrasonography of pulmonary complications in preterm infants with respiratory distress syndrome. Free supporting publication here

In this paper 120 patients with respiratory distress syndrome had U/S performed in the first 24 hours of life and then follow-up as indicated. Of this sample, 47 had complications including hemorrhage, pneumothorax, pneumonia, atelectasis and bronchopulmonary dysplasia. Of this group 45 were detected by ultrasound with the only two failures being small pneumothoraces. There were 512 ultrasounds performed compared to 612 clinically indicated chest x-rays during the same period. Again the potential reduction in ionizing radiation is astounding!

To be clear I am not suggesting that we do away with the chest x-ray but rather there is a great potential for reducing the need for the urgent chest x-ray when patient status changes. Depending on the situation in the unit you work in, an x-ray from the time it is ordered till the time it is performed and processed can be 15-30 minutes or more. The time it takes to perform a bedside ultrasound exam is 2-3 minutes at most making this modality at least in my mind a great first line option.

As I mentioned earlier though I am not an expert but merely an enthusiastic Neonatologist. This is not meant to replace the Pediatric Radiologist and in fact I would like to stress that I believe it is best suited to questions in which a binary yes/no answer is being asked. Does the patient have a pneumothorax, atelectasis, fluid in the abdomen or chest; all yes/no. Does the patient have a tumour in the liver and if so what type? That is a different kettle of fish and deserving of a Pediatric Radiologist’s opinion.

How do we go from “good to great” and utilize this technology accurately and safely. Clearly we are in need of practice, which can only come through training by qualified people who have demonstrated proficiency in the field. Thankfully in our situation we have access to individuals who have taught courses internationally and are willing to provide the training locally to us.   Our strategy will include monthly U/S rounds supervised by these two experts who will provide a small group of us “superusers” with the training and ultimately confidence to put this into routine clinical practice. As confidence rises, the use of POC U/S to obtain vascular access will develop as will performance of such procedures as a bladder tap under direct visualization to improve our success rate. I suspect new uses will develop over time as we are at the forefront of this technology. There are courses offered abroad but like the paucity of Neonatal research in the area there are very few if any such programs for the Neonate specifically.

What I can say for sure is that the benefits of POC U/S are vast and I believe our patients will benefit both during their hospital stay and into adulthood due to the reduction in ionizing radiation that I see coming for these fragile infants. I for one am excited and energized by the future of this field and look forward to an expanding body of literature on the topic.